AudiologyOnline Phone: 800-753-2160


ReSound Nexia - February 2024

How Does ReSound Aventa Prescribe MPO, and What is the Rationale?

Jennifer Schumacher, AuD, Jennifer Groth, MA

August 22, 2016

Share:

Question

How does ReSound Aventa prescribe MPO and what is the rationale?

Answer

Historically, maximum power output (MPO) values visible in the Fit screen of the ReSound Aventa fitting software reflected maximum output for the hearing aid when set to full on gain. MPO values were not prescribed for the individual because the ReSound fast-acting WDRC minimizes the risk that uncomfortable loudness levels will be exceeded. With this type of system, the hearing aid may never produce uncomfortable sounds for any input level that the user is likely to encounter (Dillon, 2012). However, displaying the MPO values reflective of full on gain settings could be misleading, as the clinician saw MPO values that were not indicative of the output levels to which the patient might actually be exposed. In addition, the introduction of linear and semilinear amplification modes, as well as slower time constant options, increased the risk of patients experiencing louder levels. The strategy for MPO settings in Aventa software were thus updated for releases of Aventa 3.8 and later.

In order to make this feature more user-friendly and to take new amplification related features into account, Aventa now displays the MPO prescription specific to each patient’s fitting. The MPO values that will be seen by the clinician in the software are impacted by several factors, including the patient’s hearing thresholds, the hearing instrument being fitted, the coupler view used in the software, and the selected fitting prescription.

Let’s begin with an example. The following hearing thresholds were entered into a patient profile in NOAH (Figure 1).

Audiogram in NOAH

Figure 1. Audiogram in NOAH.

Fittings were created using this patient profile in Aventa 3.7 (Figure 2), and in a later version of Aventa (Figure 3). The hearing thresholds, insertion gain and hearing instruments were identical in both fittings. However, the MPO values differ between the software versions. This difference reflects changing the setting from maximum output for full on gain (v3.7) to a maximum output prescription for the individual (v3.8 and later).

Hearing instrument Fit screen in Aventa 3.7

Figure 2. Hearing instrument Fit screen in Aventa 3.7. Note the MPO values reflect maximum output at full on gain settings. Therefore, they vary depending on the specific device being fit.

 Hearing instrument Fit screen in Aventa 3.8 and later

Figure 3. Hearing instrument Fit screen in Aventa 3.8 and later. Note the MPO values reflect prescribed maximum output for the individual patient based on hearing threshold levels.

How do different fitting factors influence the MPO prescription?

1. Hearing thresholds. The MPO prescription in Aventa is calculated using the NAL-NL2 formula, which is based on the patient’s hearing threshold levels. Different hearing thresholds will result in different prescribed MPO values.

2. Hearing instrument. Depending on the hearing instrument used in the fitting, the MPO values viewed in the software may differ from the MPO prescription. If the prescribed MPO values are greater than the maximum output of the hearing instrument receiver, Aventa will display the lower MPO values constrained by the instrument receiver. This may be observed when the patient’s hearing thresholds are near the limits of the fitting range at one or more frequencies.

3. Coupler and stimulus view. The MPO values visible in the software will change depending on the coupler view chosen in the settings. As Coupler view is changed from insertion gain to 2cc coupler, for example, the displayed gain and MPO values will change accordingly. The assumed verification signal for the MPO is, however, always a pure tone at 90 dB SPL regardless of the selected view. This means that if “Speech-weighted” is selected instead of “Pure tone”, then the displayed gain values will change but the MPO values will not.

4. Individual UCL. The way Aventa handles individually measured UCL values has not changed from previous versions of the software. If UCLs are measured for a patient and entered into the software, they will influence gain for 80dB inputs when the proprietary Audiogram+ fitting rule is selected. The MPO prescription will not change, but output for loud sounds will be affected by the change in G80 gain. For third-party prescriptions, whether or not individually measured UCL is taken into account depends on the particular formula. NAL-NL2 (which also provides the MPO prescription when Audiogram+ is selected) does not take individually measured UCL into account when prescribing MPO. See Dillon (2012)1 for a discussion of the rationale for this.

5. Fitting prescription. MPO values are prescribed for each patient based on the NAL-NL2 formula, unless the selected fitting formula contains its own MPO prescription. In that case, the MPO values for the selected prescription are used. This would include DSLv5 for adults and for pediatrics.

The MPO prescribed values are not targets. They are levels that the output should never exceed in order to ensure patient comfort. The actual output of the hearing aid for any given input level still depends on the gain. Therefore, when verifying MPO, it will be common to see output levels that are well under the prescribed levels.

A remaining question that frequently arises is whether to lower the MPO when the patient says the sound is too loud. Clinicians are rightly concerned about ensuring the safety and comfort of their patients in the presence of loud sounds, and it is common only to focus on the acceptable maximum settings. It might be tempting to lower MPOs from the prescribed level whenever the patient complains of the sound being too loud. However, it is also important to remember that the MPO should not be set too low, and that overall loudness complaints often are due to the gain settings rather than the MPO. The MPO is not like the feature in music players where it is possible to limit the volume to avoid hearing damage. Output limiting uses either a very high amount of compression, peak clipping, or some combination of the two to ensure that sounds above a certain level are not emitted by the receiver. This is very different than limiting volume (gain) in that it necessarily distorts the output signal when the limit is reached. This is almost always undesirable so it is also a goal of the MPO prescription to prevent the hearing aid from operating in saturation to the degree possible. Just as there is an acceptable maximum power output level, there is also a minimum acceptable power output level that will maintain good sound quality for speech and other everyday sounds. In validating the National Acoustic Laboratory’s approach to prescribing MPO, Storey and colleagues (1998) investigated the range of acceptable settings for MPO and found that it was influenced by the type of sound as well as the type of output limiting. Generally speaking, the upper limits – or the highest acceptable values the MPO could be set to – were not different for peak clipping versus compression limiting, but higher settings were accepted for impulse sounds than speech. Minimum levels were most affected by limiting type, with lower levels being acceptable for compression limiting. Based on this data as well as clinical experience, ReSound recommends keeping the channel specific MPO settings at 95 dB SPL or above if possible. The more severe the hearing loss, the trickier it becomes to balance audibility and comfort while avoiding too-frequent saturation of the hearing aid. The sound quality complaints that can lead the clinician to suspect too low MPO levels are “choppy” sound quality and sometimes the perception of cutting out by the hearing aid.

References

Dillon, H. (2012). Hearing aids. Thieme: New York.

Storey, L., Dillon, H., Yeend, I., & Wigney, D. (1998). The National Acoustic Laboratory’s procedure for selecting the saturation sound pressure level of hearing aids: Experimental validation. Ear & Hearing, 19(7), 267-279.


jennifer schumacher

Jennifer Schumacher, AuD

Audiologist, GN ReSound Global Audiology

Jennifer Schumacher, AuD is an audiologist in the GN ReSound Global Audiology group. She is involved in designing and conducting clinical research trials, analyzing data, supervising AuD students and presenting audiology information to students and clinicians. Her primary areas of interest include hearing aid technology, and using this technology to assist in providing patient-centered solutions.


jennifer groth

Jennifer Groth, MA

Director, Audiology Communications in Global Audiology

Jennifer Groth, MA, serves as Director, Audiology Communications in Global Audiology for GN ReSound Group. Jennifer’s history with GN Resound includes roles as senior research audiologist as well as product manager. Previous to joining GN ReSound, she worked as a clinical audiologist at the Gentofte University Hospital Department of Audiology in Denmark, and served as coordinator of an OAE-based neonatal hearing screening project for Copenhagen County. She holds a Master of Arts degree in Speech Pathology & Audiology from the University of Iowa.


Related Courses

The Power of the Demo: An Innovative Field Study Offers a New Perspective
Presented by Jennifer Schumacher, AuD, Meredith Buemi, AuD, Jennifer Groth, MA
Text/Transcript
ReSound

Presenters

Jennifer Schumacher, AuDMeredith Buemi, AuDJennifer Groth, MA
Course: #33574Level: Intermediate1 Hour
  'The data presented was encouraging to increase hearing aid fittings, however, the time needed to complete these demos is not feasible on a daily basis'   Read Reviews
Hearing care professionals can gain valuable insights to what their patients’ initial listening experiences might be like simply by listening to the hearing aids. In a novel experiment designed to put Hearing Care Professionals in the shoes of their patients and to illustrate how sound quality can positively color the initial experience trying hearing aids, ReSound field representatives were trained to demo ReSound LiNX Quattro hearing aids on HCPs in a similar way that HCPs might do with their patients.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

ReSound OMNIA: Introducing the Full Family
Presented by Neil Wright, AuD
Recorded Webinar
ReSound

Presenter

Neil Wright, AuD
Course: #38667Level: Introductory1 Hour
  'better understanding different directional modes'   Read Reviews
This course will introduce the expanded line of the ReSound OMNIA family of hearing aids, including the new mini-rechargeable Receiver In the Ear, as well as the updated BTE and custom models. The presentation will highlight the benefits of the OMNIA technology, and discuss changes made to ReSound Smart Fit fitting software that are introduced as part of the update to version 1.16.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

ReSound Government Services: November 2021 What’s on Contract?
Presented by Kara Sinner, MS, CCC-A, FAAA
Recorded Webinar
ReSound

Presenter

Kara Sinner, MS, CCC-A, FAAA
Course: #37135Level: Introductory1 Hour
  'The manner in which the information was presented was very clear'   Read Reviews
This presentation will review ReSound products and features available for Government Service Audiologists. Areas of discussion include ReSound hearing instruments, technology features, and ReSound Smart Fit fitting software.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

Bridging the Gap: Bimodal Fitting Considerations for Hearing Aid Clinicians
Presented by Neil Wright, AuD
Recorded Webinar
ReSound

Presenter

Neil Wright, AuD
Course: #33557Level: Intermediate1 Hour
  'Comprehensive presentation of the complex relationship within a bimodal fitting'   Read Reviews
Research has shown the benefits of bimodal stimulation in structured clinical trials, but clinical patients may require programming outside of the standard fitting flow. This presentation will discuss bimodal fitting considerations for hearing aid fitting clinicians to help patients get the most of their bimodal stimulation.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

Fitting Assistance at a Distance: ReSound Live Assistance for the Government Services Clinician
Presented by Kara Sinner, MS, CCC-A, FAAA
Recorded Webinar
ReSound

Presenter

Kara Sinner, MS, CCC-A, FAAA
Course: #35099Level: Introductory1 Hour
  'It was a good introduction to remote programming'   Read Reviews
This presentation will introduce the process for performing remote fine-tuning sessions using ReSound Assist: Live Assistance. Details will describe steps as they pertain to both the patient and the Government Services clinician.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

Our site uses cookies to improve your experience. By using our site, you agree to our Privacy Policy.